Preparations for use S88 / OPMI Lumera T
Version 8.0
Page 162 G-30-1682-en
Preparing the device for sterile use
Fit asepsis caps
The device can be equipped with resterilizable products for sterile use.
The included ZEISS Asepsis Sets contain Asepsis Caps and handles which can
be sterilized in autoclaves. For more detailed information about sterilization,
please refer to the enclosed "Preparation of resterilizable products" manual
for each Asepsis Set.
The following asepsis caps can be attached:
1 sterilizable caps, 49 mm (180° tiltable tube)
When using the 180° tiltable tube, the asepsis cap is used for sterile
operation of the wheel used to set the interpupillary distance.
2 sterilizable handgrips (OPMI VISU)
3 sterilizable caps 22 mm
for operating buttons with a diameter of 22 mm
CAUTION Risk of infection from contaminated asepsis accessories!
• Use only sterile asepsis accessories suitable for the system.
• You must clean, disinfect and sterilize asepsis accessories before each
use. This also applies to the first use after delivery.
• Make sure that the operating personnel have been instructed in the
sterile operation of the device.
• Before each use of the surgical microscope, check to ensure that the
asepsis caps are firmly attached to the surgical microscope.