Version 8.0
G-30-1682-en Page 161
S88 / OPMI Lumera T Preparations for use
Æ The microscope is now adjusted so that a constantly sharp image is
produced throughout the entire magnification range without your having
to refocus after changing the magnification value.
If this is not the case, repeat the procedure.
Adjusting the eyecups
• Adjust the eyecups so that you can survey the entire field of view.
– Viewing with eyeglasses: screw in the eyecups.
– Viewing without eyeglasses: remove the eyecups.
Adjusting the working distance and magnification
• Position the microscope above the surgical field so as to ensure
convenient work.
• Focus roughly on the object by changing the distance between
microscope and object.
• Set the microscope to the highest magnification and correct the fine
focusing until the image is defined sharply.
• Set the microscope to the required magnification. The focal plane is
retained.
CAUTION Risk of injury to the patient caused by parts falling down!
Due to collisions or incorrect installation, parts may fall down and injure the
patient.
• Before using the device, check that all parts are properly and securely
mounted.
• Check that the floor stand is secured in a steady position and can be
pivoted without causing collisions.