S88 / OPMI Lumera T Safety measures
• In the event that the zoom function can no longer be controlled via the
foot control panel or the control panel, adjust the zoom factor manually
at the microscope.
• In the event that the focusing function can no longer be controlled via the
foot control panel or the control panel, use the suspension arm to focus
the surgical field. This may cause injury to the patient or damage to
mounted accessories. Ensure to provide sufficient clearance between
equipment and patient (min. 40
mm) when focusing manually. Ensure
that attached accessories may not come into contact with other objects.
• In the event that the X-Y movement of the surgical microscope can no
longer be controlled via the foot control panel or the control panel,
position the device manually via the suspension arm. Again, you should
ensure to provide sufficient clearance between equipment and patient
(min. 40
mm).
• In the event that the light intensity can no longer be controlled via the foot
control panel or the control panel, regulate the illumination manually at
the suspension system.
• If there is a failure and you cannot get any help from the chapter "What
to do in the event of malfunctions", attach a sign "Out of order" to the
system and contact ZEISS service representative.
• Do not pull at the light guide, power cord or any other connecting cables.
• Constantly monitor the system during use.
• Risk of crushing - mind your fingers!
Fingers may be crushed in the areas marked with the "Risk of crushing" sign.
Do not touch these areas while the device is being moved.