What to do in the event of malfunctions S88 / OPMI Lumera T
Version 8.0
Page 210 G-30-1682-en
Insufficient surgical field
illumination.
Brightness level set too low. Adjust brightness on
suspension system's display
field or using foot control
panel.
Aged xenon lamp. Switch off light source
on stand.
Use an OR lamp to
illuminate the surgical field.
Change xenon lamp module.
Page 196
Defective light guide
(illumination not uniform).
Notify Service.
The light guide probably
needs to be replaced.
Surgical field illumination
too bright.
Brightness level set too
high.
Adjust brightness on
suspension system or foot
control panel.
Page 156
Switch off light source
on stand.
Use an OR lamp to illuminate
the surgical field.
Notify Service.
No surgical field
illumination.
Xenon lamp does not ignite. Switch off light source
on stand.
Use an OR lamp to illuminate
the surgical field.
Notify Service.
Xenon lamp is lit, but beep
sounds intermittently.
Defective lamp control
system.
Switch off light source
on stand.
Use an OR lamp to illuminate
the surgical field.
Notify Service.
Surgical field illumination on
microscope not working.
Thermal cut-out activated. Remove the cause of
overheating. For example,
drapes could be covering
the screen. When the lamp
module has cooled down,
the light source switches on
again.
Fault Possible reason Remedy Cross
reference