Safety measures S88 / OPMI Lumera T
Version 8.0
Page 32 G-30-1682-en
Superlux Eye light source with integrated halogen light source (option)
• For the lamp change procedure and how to reset the service hour counter,
see "Switching to the backup lamp" on
page 90.
1 Flap
The flap is used as the mechanical indicator for the operating status of the
halogen lamps.
– When the flap is closed, the main lamp is operative.
– When the flap is open, the main lamp has failed. The backup lamp is on.
2 Switching to the backup lamp
The lamp housing contains a backup lamp which automatically swings
into position when the first lamp fails. If this automatic function fails,
you can switch on the backup lamp by pressing this button.
3 Filter selector
The filter selector knob has four positions:
4 Indicator lamp (color: amber)
– is lit when the main lamp has failed. In addition, open flap (1) on the
lamp module indicates that the main lamp has failed. The backup lamp
is on.
CAUTION The xenon lamp has a limited service life of 500 h!
If used beyond its maximum service life, the xenon lamp may explode.
• Please replace the xenon lamp in due time.
• Reset the service hour counter to "0" after replacing the lamp.
No filter
blue barrier filter (retina protection filter): use the blue barrier
filter when operating on the eye. It protects the patient's
retina against unnecessary (blue) radiation and increases the
recommended exposure time by factor 3.
KK 40 filter:
to increase the color temperature
Empty filter position