Safety measures S88 / OPMI Lumera T
Version 8.0
Page 34 G-30-1682-en
Manual mode
1 MANUAL key
The Manual button permits you to switch to manual mode. The motorized
functions of the surgical microscope are deactivated. The lamp brightness
is automatically adjusted to a default setting. This lamp brightness value is
displayed in the first display field.
When you have switched to the manual mode, the amber LED is lit and
the blinking text "MANUAL" appears in the third display field
The surgical microscope can no longer be operated via the foot control
panel, the handgrips or the display and control panel.
In the manual mode, you can use the foot control panel only to switch
the illumination on and off, and you can unlock the magnetic brakes by
pressing the button on the microscope.
The manual mode is retained even if you switch the system off and back
on at the power switch.
Press the Manual button a second time to reactivate electronic control.
The basic mode is displayed again on the display and control panel.