Troubleshooting 8-5
Note: The next step is irreversible. All settings including custom reference
ranges and the date and time —are returned to their factory values, and
the memory is cleared. Results previously stored, however, are not
affected.
4. Press CONTINUE to restore all factory settings (including the
clock) and clear all memory from the analyzer.
5. Refer to the sections noted below to reset the date and time, any custom reference ranges,
and the units of measurement:
“Customizing Reference Ranges” on page 5-2
“Changing Date and Time” on page 5-14
“Selecting Units” on page 5-16