3-2 Set Up and Analyzer Description
3.2 Setting Up the Analyzer
Refer to the required environmental specifications in Section 3.
1. Check that the VetScan VS2 is:
On a sturdy, level surface, free of vibration and sudden jolts.
In an ambient operating temperature of 15–32 ºC (59–90 ºF).
In an environment free of animal hair, dust, and other contaminants.
Not placed near a sunny window or another heat source.
At least six inches from the wall for access to the power connection and USB
ports.
Note: If the analyzer is subjected to an electrostatic discharge event, you may
need to restart the unit.
2. Plug the power jack into the analyzer, and
plug the detachable power supply cord into
the power adapter and into a grounded elec-
trical outlet.
Note: To prevent power surges or drain,
do not plug the analyzer into the
same circuit as a centrifuge or
any other high-current device. If
this is not possible, use an ancil-
lary surge protector or battery
backup for the analyzer.
WARNING: Use only the VS2 power supply. Any other power supply will damage the
instrument and void the warranty.