Maintenance and Service 10-3
Changing an Air Filter With a Cover
1. Turn the analyzer off by pressing and holding the Power button.
2. Remove the power cord from the back of the analyzer.
3. Remove the filter cover from the back
of the analyzer.
4. Remove the filter.
5. Wash the filter with warm water to
remove dust, animal hair, and any
other debris.
6. Dry the filter and place it back into
the analyzer.
CAUTION: Make sure the filter is
completely dry before
replacing it.
7. Snap the filter cover back into place
over the filter.
8. Plug the power cord into the back of the analyzer.
9. Press the Power button. The analyzer then performs a diagnostic self test, and may take a
few minutes to bring the rotor chamber to operating temperature.