Test Procedure and Results 4-5
4.4 Running Tests
This section includes detailed, step-by-step instructions for performing analyses using the VS2.
Note: The VS2 includes a number of optional data functions that can be used
to enter more detailed information about samples.
In this manual, the screens for these functions are marked with this
icon: .
However, the screens for these functions will appear on the analyzer
only if the functions are enabled. If the screen shown in the manual
does not appear on the analyzer screen, skip that step.
See page 5-30 for details about these functions.
1. Turn on the analyzer power by pressing the Power button on the front of the analyzer.
Do this before removing reagent rotors from the refrigerator.
The analyzer starts up, then
performs a self test.
See Section 7.2, “Quality
Control Features” for infor-
mation about the analyzer
self test. Alternatively, the
analyzer can be left on.
The analyzer’s heaters may require time to warm the analyzer
rotor chamber to operating temperature. If so, the analyzer dis-
plays this while warming up:
Performing iQC...
VetScan VS2 V1.2.3.4