Troubleshooting 8-3
These symbols (“flags”) are printed instead of a result:
To obtain more information for results with (H) HEM, (L) LIP or (I) ICT, print an error
report (see Section 6). The error report shows a value with HEM, LIP or ICT next to it.
The entire numerical line displayed on the error report, including the line HEM, LIP, and ICT,
must be interpreted by an Abaxis Technical Service Representative to determine whether results
are usable. Call Abaxis Technical Service at 1-800-822-2947.
8.4 Out-of-Range Samples
Out-of-range samples are indicated in the printed results. See Section 4.6, “Using Test Results”
for an explanation of the various symbols printed in the results.
An out-of-range sample may be able to be diluted. See Section 8.6, “Sample Dilution” for instruc-
tions. A sample that is found to be out-of-range more than once should be tested by an alternate
laboratory-approved method, or sent to a referral laboratory.
8.5 Rotor Cancellations and Procedures
If the rotor cancels, record the following information or print an error report, and contact Abaxis
Technical Service at 1-800-822-2947:
Lot number
Product name
4-digit error code
Specimen type and species
Chemistry interference
Hemolysis interference
Lipemia interference
Icterus interference
~~~
HEM
LIP
ICT