8-4 Troubleshooting
8.6 Sample Dilution
The following customer-developed protocol can be used for diluting samples.
Dilution of Sample 1:1 with Saline
(Use serum or heparinized plasma)
1. Combine one part (100 PL) serum or heparinized plasma with one part (100 PL) normal
saline (0.9%) (do not use distilled water) into a clean, red top tube and gently mix. Use the
diluted sample for testing.
2. Run the diluted sample with a new rotor.
3. Print the results.
4. Multiply the values in the results by 2.
8.7 Reinitializing the Analyzer
The analyzer may require reinitialization if its memory becomes corrupted. Reinitializing returns
all settings to their original factory default values, including the clock, and also clears all patient
results, control results, and custom reference ranges from memory. Perform the following steps to
reinitialize the analyzer.
1. In the Home screen, press the Settings icon.
2. Press the More Settings icon.
3. Press the Factory Settings icon.
Note: The analyzer does not provide accurate values for albumin, calcium,
globulins, or sodium. Diluting the sample can introduce error into the
final result.