4-12 Test Procedure and Results
CAUTION: Dispose of the rotor according to the lab’s standard procedures for animal
samples.
24. To analyze another sample, insert a new rotor and repeat the above procedure.
25. When finished, press CLOSE to close the drawer and return the
analyzer to standby mode.
4.5 Canceling an Analysis
1. To cancel an analysis in process, press CANCEL on the touch-
screen.
2. Press CANCEL to con-
firm. The analysis is
then cancelled, and the
rotor drawer opens.
3. Remove the rotor from the drawer. The analyzer is now ready to perform another analysis.
Press CLOSE to close the drawer.
Confirm
NOCANCEL
Cancel Analysis?
Remove Rotor
CLOSE
To analyze additional sam ple,
load new rotor