Configuring the Analyzer 5-3
4. Select the reference range to modify.
■ To modify the last reference range that was changed,
press Last Modified, then skip to step 4 on page 5-4.
■ To modify any reference range, press All, then continue
to step 5.
5. The Modify Reference Ranges screen then opens. From here,
the following procedures are available:
■ Modify the reference range for a specific analyte:
see below.
■ Modify a set of ranges associated with a particular spe-
cies: see page 5-4.
■ Add a reference range for a new species: see page 5-5.
■ Delete a species: see page 5-9.
Modifying Reference Ranges of a Single Analyte
Change the reference ranges for a specific analyte as follows.
1. In the Modify Reference Ranges screen, press Analyte.
2. Select the analyte from the list. Use the up (S) and down (T)
arrow keys to scroll through the list.
Include Reference Ranges
Last Modified: ALB
All
Back Home
Modify Reference Ranges
Analyte
Species
Back Home
AnalyteList
HomeBack
ALB
ALP
ALT
AMY
AST