General Information 2-7
■ Can I back up the results on my VS2?
No, but you can print out reference ranges and results, or transfer them for storage in your
lab management software.
■ Can patients other than equine, large animal, or avian use the Equine, Large Ani-
mal, and Avian profiles?
No. Profiles and reference ranges are designed for the specific species indicated on the
package label. Running samples of other species can provide inaccurate results because
some analytes are species-specific. See the rotor package insert for more information.
■ My results do not match clinical observation. Does my VS2 need service?
If your results are extremely low, and would not seem to support life (for example,
Ca < 4.0), this is likely the result of too little sample. This prevents the rotor from distrib-
uting the sample correctly, resulting in extremely low results for more than three chemis-
tries. Rerun the sample for accurate results.