USER MANUAL | ICOS | INSTRUCTIONS | UM/ICOS-EN REV. B.2
42
Table 8 describes fault criteria for the Temperature Alarms.
Table 8: Fault Criteria for Temperature Alarms
Status
Sensor Read
Fault Condition
Description
Analyzer Temp
Temperature High/Low
Alarm
The temperature exceeds the operating
temperature range.
11
CROSS OVER
14
Analyzer Temp
Temperature High/Low
Warning
The temperature is > the high warning
set point.
The temperature is < the low warning
set point.
17
Fault
NaN reading
Occurs when there is a false or
undefined value. (NaN= not a number)
19
Dead Band
20
Acceptable
Range
No warning/alarm
No warning/alarm
If the Alarm Status is Yellow or Red,please refer to the Maintenance section
on page 72. If issue continues, please contact icos.support@ca.abb.com.