6146/6156 DC Voltage/Current Generators Operation Manual
3.2.14 STEP TIME/SEL Key
3-12
3.2.14 STEP TIME/SEL Key
Sets the step time of the program function.
The step time can be set in the range from 20 ms to 10 s by 10 ms.
3.2.15 TRIG HOLD Key (Trigger and Hold for Memory Recall/Sweep Source)
TRIG key: Functions as source trigger key.
Program mode
Memory recall mode/sweep source mode
Memory recall mode
Hold scan Pressing the TRIG HOLD key executes scan by one step.
Single scan Pressing the TRIG HOLD key executes scan from the first
number to the last number.
Pressing the TRIG HOLD key while the scan is running pauses
the scan. Pressing the TRIG HOLD key again resumes the scan.
Repeat scan Pressing the TRIG HOLD key executes scan repeatedly from the
first number to the last number.
Pressing the TRIG HOLD key while the scan is running pauses
the scan. Pressing the TRIG HOLD key again resumes the scan.
Sweep source mode
Hold sweep Pressing the TRIG HOLD key executes sweep by one step.
Single sweep Pressing the TRIG HOLD key executes scan from the start value
to the stop value.
Pressing the TRIG HOLD key while the sweep is running pauses
the sweep. Pressing the TRIG HOLD key again resumes the
sweep.
Repeat sweep Pressing the TRIG HOLD key executes sweep repeatedly from
the start value to the stop value.
Pressing the TRIG HOLD key while the sweep is running pauses
the sweep. Pressing the TRIG HOLD key again resumes the
sweep.

Table of Contents