EasyManua.ls Logo

ADCMT 6146 - Page 147

ADCMT 6146
260 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6146/6156 DC Voltage/Current Generators Operation Manual
4.3 Compatibility
4-35
4.3 Compatibility
This section describes the compatibility with the former model, 6144.
1. The 6146/6156 is equipped with the 6144-compatible mode. Set this mode using MENU and I/F.
For more information, refer to Section 5.1.2, "Compatibility" and Section 5.8.4, "6144-compatible Re-
mote Command List."
2. There is no 10 V range. The 30 V range is provided instead.
3. As the settling time is changed from 50 ms to 10 ms, the 6146/6156 may not reproduce the output of the
6144 depending on the output waiting time.

Table of Contents

Related product manuals