6146/6156 DC Voltage/Current Generators Operation Manual
4.3 Compatibility
4-35
4.3 Compatibility
This section describes the compatibility with the former model, 6144.
1. The 6146/6156 is equipped with the 6144-compatible mode. Set this mode using MENU and I/F.
For more information, refer to Section 5.1.2, "Compatibility" and Section 5.8.4, "6144-compatible Re-
mote Command List."
2. There is no 10 V range. The 30 V range is provided instead.
3. As the settling time is changed from 50 ms to 10 ms, the 6146/6156 may not reproduce the output of the
6144 depending on the output waiting time.