6146/6156 DC Voltage/Current Generators Operation Manual
5.8.3 Remote Command List
5-31
Source
Selection of
thermocouple
standard (6156
only)
JSV0 Thermocouple JIS standard C1602-1995 (new)
JSV1 Thermocouple JIS standard C1602-1981 (old)
When TC8 (type N) is specified, JSV0 will be applied
even if JIS2 (JIS C1602-1981) is selected.
JSV? Query of the thermocouple standard
Response: JSV0 or JSV1
Voltage source
value
SOV ±data [unit]
Voltage source value setting
*1
For the fixed range setting
SOV±ddddd: fixed range
SOV±dddddunit: auto range
For the Auto range setting
SOV±ddddd: auto range
SOV±dddddunit: auto range
Unit V: volt
MV:millivolt
UV: microvolt
Example when setting -1.5 mV
SOV-1.5E-3
SOV-15.MV
• Numeric values below the lowest digit in each range
are rounded off.
• For the fixed range setting, out-of-range setting is
not allowed.
• The source value cannot be set during thermal EMF
output.
0
SOV?
Query of the voltage source value
*2
Response: SOV±d.ddddE±d
Current source
value
SOI ±data [unit]
Current source value setting
*1
For the fixed range setting
SOI±ddddd: fixed range
SOI±dddddunit: auto range
For the Auto range setting
SOI±ddddd: auto range
SOI±dddddunit: auto range
Unit A: ampere
MA:milliampere
UV: microampere
Example when setting -1.5 mA
SOI-1.5E-3
SOI-15.MA
• Numeric values below the lowest digit in each range
are rounded off.
• For the fixed range setting, out-of-range setting is
not allowed.
• The source value cannot be set during thermal EMF output.
0
SOI?
Query of the current source value
*2
Response: SOI±d.ddddE±d
*1: The setting range of the source value varies depending on the model, range or function. For more information, refer to Chapter 8, "SPEC-
IFICATIONS."
*2: The number of digit and the position of the decimal point vary depending on the model, range or function. For more information, refer to
Section 5.7, "Source Value Output Format (Talker Format)."
Class Item Command Description
Default Operation
Power
ON
Factory
shipment
Memory Sweep