6146/6156 DC Voltage/Current Generators Operation Manual
5.8.3 Remote Command List
5-32
Source
Thermal EMF
source value
(6156 only)
TMP ±data
Thermal EMF source value setting
*1 23
TMP? Query of the thermal EMF source value
Response: TMP±dddd.dE±d
Room tempera-
ture compensa-
tion (6156 only)
TCO0 Room temperature compensation OFF
TCO1 Room temperature compensation ON
TCO? Query of the room temperature compensation
Response: TCO0 or TCO1
TMC ±data Room temperature setting
data: -25.0 to +85.0 (unit: °C)
0
TMC? Query of the room temperature setting value
Response: TMC±dddd.dE±d
Limit value LMV data1[,data2] Voltage limit value setting
• Setting range: ±1.0 to ±32.0
• When data2 is omitted, the same limit value as data
1 with the reverse polarity is set.
• Specify different polarities for data1 and data2
respectively.
• The negative polarity data is set as LO limit value
and the positive polarity data is set as HI limit value.
±32V
()
LMV? Query of the voltage limit value
Response: LMV<-d.ddE±d>,<+d.ddE±d>
LMI data1[,data2] Current limit value setting
• Setting range: ±0.005 to ±0.220
• When data2 is omitted, the same limit value as data
1 with the reverse polarity is set.
• Specify different polarities for data1 and data2
respectively.
• The negative polarity data is set as LO limit value
and the positive polarity data is set as HI limit value.
±220
mA
()
LMI? Query of the current limit value
Response: LMI<-d.ddE±d>,<+d.ddE±d>
Remote sense RS0 2 Wire
()
RS1 4 Wire
RS? Query of the remote sense
Response: RS0 or RS1
Count fix at
range change
CFIX0 The count value is not fixed at the range change.
CFIX1 The count value is fixed at the range change.
CFIX? Query of the count fix at the range change
Response: CFIX0 or CFIX1
*1: The setting range of the source value varies depending on the model, range or function. For more information, refer to Chapter 8, "SPEC-
IFICATIONS."
Class Item Command Description
Default Operation
Power
ON
Factory
shipment
Memory Sweep