6146/6156 DC Voltage/Current Generators Operation Manual
5.8.3 Remote Command List
5-33
Source
Continuous vari-
able digit setting
CUP0 E+0 digit continuous increase
CUP1 E+1 digit continuous increase
CUP2 E+2 digit continuous increase
CUP3 E+3 digit continuous increase
CUP4 E+4 digit continuous increase
CUP5 E+5 digit continuous increase (6156 only) (VS/IS only)
CDN0 E+0 digit continuous decrease
CDN1 E+1 digit continuous decrease
CDN2 E+2 digit continuous decrease
CDN3 E+3 digit continuous decrease
CDN4 E+4 digit continuous decrease
CDN5 E+5 digit continuous decrease (6156 only) (VS/IS only)
Polarity mode PLM0 Bipolar polarity mode
PLM1 Unipolar polarity mode
PLM? Query of the polarity mode
Response: PLM0 or PLM1
Standby mode BY0 Standby mode 0
BY1 Standby mode 1
BY? Query of the Standby mode
Response: BY0 or BY1
Query of the
source value
SRC?
Responds with the voltage or current source value.
*2
(For thermal electromotive force, returns its equivalent
voltage source value.)
Response:
SOV±d.ddddE±d
or
SOI±d.ddddE±d
*2: The number of digit and the position of the decimal point vary depending on the model, range or function. For more information, refer to
Section 5.7, "Source Value Output Format (Talker Format)."
Class Item Command Description
Default Operation
Power
ON
Factory
shipment
Memory Sweep