6146/6156 DC Voltage/Current Generators Operation Manual
5.8.3 Remote Command List
5-34
Memory
Memory setting Nnnn
P
Voltage/current
*1
1. Nnnn,: memory number 0 to 4999
2. SVRn, or SIRn, : source range
Auto range setting when omitted
3. SOV±data, or SOI±data,
: voltage/current source value
4. Nnnn: next memory number
Auto increment when omitted
5. SOV±data, or SOI±data,: next source value
~ ~
6. P: memory setting mode complete
Thermal electromotive force
1. Nnnn,: memory number 0 to 4999
2. TCn : type of thermocouple
Type T thermocouple selected when omitted
3. JSVn,: thermocouple standard
C1602-1995 selected when omitted
4. TMP±data,:
: thermal electromotive force source value
5. Nnnn: next memory number
Auto increment when omitted
6. TMP±data,: next source value
~ ~
7. P: memory setting mode complete
• Voltage, current and thermal electromotive force can
be mixed.
N0 30
mV
range
0
*5
N?nnn
Query of the specified memory number
*2
Response:
N<nnn>,SVR<n>,SOV<±data>,P
or N<nnn>,SIR<n>,SOI<±data>,P
or N<nnn>,TC<n>,JSV<n>,TMP<±data>,P
NP? Query of the memory setting status
Response:
0 memory setting complete
1 memory setting in progress
0
RSAV Saves the memory data.
RLOD Loads the memory data.
RCLR Initializes the memory data.
• Sets Standby.
• Saved data is not initialized.
*1: The setting range of the source value varies depending on the model, range or function. For more information, refer to Chapter 8, "SPEC-
IFICATIONS."
*2: The number of digit and the position of the decimal point vary depending on the model, range or function. For more information, refer to
Section 5.7, "Source Value Output Format (Talker Format)."
*5: Not initialized by the RINI or *RST command."
Class Item Command Description
Default Operation
Power
ON
Factory
shipment
Memory Sweep