6146/6156 DC Voltage/Current Generators Operation Manual
5.8.3 Remote Command List
5-35
Memory
First number
Last number
SC[mmm,]nnn First number and last number setting
SCmmm,nnn
mmm: first number 0 to 4999
nnn: last number 0 to 4999
mmmm>nnn: not allowed
The first number is set to 0 when omitted.
0,
4999
SC? Query of the first number and the last number
Response:SC<mmm>,<nnn>
mmm: first number 0 to 4999
nnn: last number 0 to 4999
Step time SPdd.dd Step time setting
SPdd.d dd.d: 0.02 to 10.00 (unit: sec)
0.1
()
SP? Query of the step time
Response: SPdd.dd
Scan mode SCM0 Hold scan mode
()
SCM1 Single scan mode
SCM2 Repeat scan mode
SCM? Query of the scan mode
Response: SCM0 to SCM2
Memory number RCLMnnn Specifies the memory number.
RCLMnnn,: memory number 0 to 4999
0
RCLM? Query of the current memory number
Response: RCLM000 to RCLM4999
Scan start *TRG Hold scan:
Increments the memory number.
Single/repeat scan:
Starts or pauses the scan.
Scan stop SWSP Stops the scan.
Sweep
Sweep value SN
±start,
Sweep data setting
*1
±start: start value
0.001m
V
0.0001
mA
23.1°C
±stop, ±stop: stop value
0.1mV
0.01m
A
33.0°C
±step ±step: step value (The polarity is ignored.)
0.001m
V
0.0001
mA
0.1°C
The optimum range that can output these values is
set when sweep is executed.
Any of these values cannot be omitted.
Values for the currently selected function are set.
SN?
Query of the sweep data
*2
Response:SN<
±start>,<±stop>,<±step>
*1: The setting range of the source value varies depending on the model, range or function. For more information, refer to Chapter 8, "SPEC-
IFICATIONS."
*2: The number of digit and the position of the decimal point vary depending on the model, range or function. For more information, refer to
Section 5.7, "Source Value Output Format (Talker Format)."
Class Item Command Description
Default Operation
Power
ON
Factory
shipment
Memory Sweep