EasyManua.ls Logo

ADCMT 6146 - Page 185

ADCMT 6146
260 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6146/6156 DC Voltage/Current Generators Operation Manual
5.8.3 Remote Command List
5-37
Remote
Block delimiter DL0 CR/LF +EOI *4
()
DL1 LF
DL2 EOI
DL3 LF +EOI
As EOI is a GPIB function, it is not output by USB.
DL? Query of the block delimiter
Response: DL0 to DL3
Service request S0 SRQ transmission is allowed. *4
()
S1 SRQ transmission is not allowed.
S? Query of the SRQ transmission
Response: S0 or S1
Status *STB? Status byte reading
Response: ddd
*SRE *6 Service Request Enable Register setting (0 to 255)
Response: ddd
Depen
ding
on
*PSC
0 *5
*SRE? *6 Response: ddd
*ESR? *6 Query of the Standard Event Status Register (ESR)
Response: ddd
*ESE *6 Standard Event Status Enable Register setting (0 to
255)
Response: ddd
Depen
ding
on
*PSC
0 *5
*ESE? *6 Response: ddd
DSR? *6 Query of the Device Event Status Register (DSR)
Response: ddddd
DSE *6 Device Event Status Enable Register setting (0 to
65535)
Response: ddddd
0 *5
DSE? *6 Response: ddd
*CLS *6 Status clear
Operation com-
plete
*OPC After all operations are complete, set LSB of the Stan-
dard Event Status Register.
*OPC? Response: 1 (after all operations completed)
*WAI Waits until all operations are complete (GPIB only).
*4: Not initialized by the RINI command.
*5: Not initialized by the RINI or *RST command.
*6: Not available in the 6144-compatible mode.
Class Item Command Description
Default Operation
Power
ON
Factory
shipment
Memory Sweep

Table of Contents

Related product manuals