6146/6156 DC Voltage/Current Generators Operation Manual
5.8.3 Remote Command List
5-38
Remote
Status clear flag
at power ON
*PSC<n> n: ±32767
n
≠0: Initializes the following registers at power ON.
n=0: Not initializes the following registers at power
ON.
•Service Request Enable Register
•Standard Event Status Enable Register
0
*PSC? Response: 0 or 1 (when setting to other than 0.)
System
Initialization C Device clear
*RST Initializes parameters.
Instrument infor-
mation
*IDN? Tells the instrument to ask itself.
Response:
ADC Corp.6146,nnnnnnnnn,xxxx
or
ADC Corp.6156,nnnnnnnnn,xxxx
nnnnnnnnn: serial number
xxxx: software revision
Self test *TST? Executes the test and reads out the results.
Response:
0; Pass
1; Fail
(Not executable during Operate)
Error log ERR? Reads out error descriptions.
Response: ±ddd, "xxxxxxxxxx"
• Records up to 20 error logs by the FIFO method.
• When 21th or later error occurs, the last saved error
is replaced by -350 "Queue overflow."
• When there is no error, returns +000 "No error."
Notice buzzer NZ0 OFF
()
NZ1 ON
NZ? Query of the notice buzzer setting
Response: NZ0 or NZ1
Limiter detec-
tion buzzer
UZ0 OFF
()
UZ1 ON
UZ? Query of the limiter detection buzzer setting
Response: UZ0 or UZ1