6146/6156 DC Voltage/Current Generators Operation Manual
8.2 Source Function
8-6
8.2 Source Function
Memory recall: Reads out a specified memory number.
Reads out memory numbers by turn at intervals specified by the step time.
Scan mode: Hold; Reads out a memory number by trigger input.
Single; Scans once from the first number through the last number.
Repeat; Scans repeatedly from the first number through the last number.
Maximum memory: 5000 data
Step time: Setting range; 0.02 to 10.00 s, setting resolution; 0.01 s
Sweep function: Increments or decrements the source value by the step value from the specified
start value to stop value.
Suspend function: Can be selected between HiZ (high resistance) and LoZ (low resistance) with
the output relay ON.
Continuous variable function:Increments or decrements the current setting value to the full scale or zero at
intervals specified by the step time.
Limiter: Voltage and current limiters are separately set.
Current limiter in voltage source
Voltage limiter in current source
HI and LO limit values can be set separately.
Output system: Floating, bipolar
Output terminal: Front Safety socket
HI OUTPUT, HI SENSE, LO OUTPUT, LO SENSE
Maximum input: HI-LO; 0.5 V peak Max (30 mV/300 mV range)
32 V peak Max (3 V/30 V range)
OUTPUT-SENSE; 0.3 V peak Max
LO-chassis; 500 V peak Max
Maximum remote sensing voltage:
HI OUTPUT - HI SENSE ; ±0.3V MAX
LO OUTPUT - LO SENSE ; ±0.3V MAX
(The voltage between HI SENSE and LO SENSE must be within the maxi-
mum output voltage range)
GPIB interface: Compliant with IEEE-488.2-1987
Interface function; SH1, AH1, T6, L3, SR1, RL1, PP0, DC1, DT1, C0, E2
Connector; Amphenol 24 pin
USB interface: USB 2.0 Full-speed
Connector; Type B
BCD parallel interface: Remote programming; output level, polarity, range, Operate, load signal
Connector; Amphenol 36 pin
NOTE: factory option