6146/6156 DC Voltage/Current Generators Operation Manual
2.4.1 Thermal Electromotive Force Output Operation
2-56
5. Pressing V/I displays the voltage equivalent to the thermal electromotive force
of the setting temperature.
The voltage source value cannot be changed in this state.
6. Pressing V/I again returns to the temperature display.
When the TC.OUT is set to ON, the output source cannot be switched to current.
To output current, set TC.OUT to OFF.
Selecting the thermocouple
The thermocouple is selected using MENU, TEMP and TC.TYP. Thermocouple type
J is selected in the following example.
1. Press MENU.
2. Select ..4.. TEMP with , or , and enter the Select level with .
3. Select TC.TYP with , or , and enter the Input/Run level with .
4. Select TC-J with , or , and press ENTER and EXIT in this order to
return to the HOME screen.
Type of thermocouple
Cursor (blinking)
Type of thermocouple
Cursor (blinking)