6146/6156 DC Voltage/Current Generators Operation Manual
2.5.2 Saving Parameters
2-60
2.5.2 Saving Parameters
2.5.3 Loading Parameters
Operation Character display area
1. 6146:
Press MENU and select ..5.. PARAM with , or .
MENU
..5.. PARAM
6156:
Press MENU and select ..6.. PARAM with , or .
MENU
..6.. PARAM
2.
Enter the Select level with .
PARAM
P.OFF PON.LD
3.
Select SAVE with , .
PARAM
SAVE
4.
Enter the Input/Run level with .
SAVE
Ent USER0
Select the save area (USER0 to USER3) with , or .
SAVE
Ent USER2
5. Press ENTER.SAVE
BUSY
6. When saving is complete, "Done" is displayed and the menu reverts to
the Select level.
SAVE
DONE
7. Press EXIT to exit the menu. PARAM
SAVE
Operation Character display area
1. 6146:
Press MENU and select ..5.. PARAM with , or .
MENU
..5.. PARAM
6156:
Press MENU and select ..6.. PARAM with , or .
MENU
..6.. PARAM
2.
Enter the Select level with .
PARAM
P.OFF PON.LD
3.
Select LOAD with , or .
PARAM
LOAD
4.
Enter the Input/Run level with .
LOAD
Ent USER0
Select the load area (USER0 to USER3/DFLT) with , or .
LOAD
Ent USER2
5. Press ENTER. LOAD
BUSY