1100 Series DAD and MWD User Manual 133
Maintenance 8
5
Take the new inlet capillary and bend it 90° about 35 mm
from its end.
6 Bend the capillary again by 90° as shown below.
7 Insert the capillary into the hole between fixing screw and
the inlet fitting.
8 The capillary lays in the grove and should be tied around
the body (in the grove) 5 times.
35 mm
90°
90°