134 1100 Series DAD and MWD User Manual
8Maintenance
9
Insert the fixing screw, so that the capillary cannot leave
the grove.
10 Fix the flow cell body to the heat exchanger.
11 Fix the inlet capillary to the flow cell body handtight first.
Then do a 1/4 turn with a 4-mm wrench.
12 Fix the new heat exchanger to the clamp unit.