1100 Series DAD and MWD User Manual 141
Maintenance 8
16
The quartz body and the cell seal assembly can be
separated for cleaning purpose.
17 This figure shows the correct holding of the quartz body
and the cell seal assembly.
18 Replace the cell seal assembly onto the quartz body.
Always use a new seal assembly to exclude damage
during disassembling.
19 Slide the quartz body completely into the cell body to the
front stop (use for example a toothpick).