1290 Infinity II High-Speed Pumps User Manual 112
7 Error Information
General Error Messages
Leak Sensor Short
Error ID: 0082
The leak sensor in the module has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak sensor current to
change within defined limits. If the current increases above the upper limit, the
error message is generated.
Probable cause Suggested actions
1 Defective leak sensor.
Please contact your Agilent service representa-
tive.
2 Leak sensor incorrectly routed, being
pinched by a metal component.
Please contact your Agilent service representa-
tive.
3 Power switch assembly defective
Please contact your Agilent service representa-
tive.
4 Cable or contact problem.
Please contact your Agilent service representa-
tive.