EasyManuals Logo

Agilent Technologies 1290 Infinity II User Manual

Agilent Technologies 1290 Infinity II
291 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
1290 Infinity II High-Speed Pumps User Manual 47
3Using the Pump
Recommendations for Bio-Compatible and Bio-Inert Systems
Recommendations for Bio-Compatible and
Bio-Inert Systems
Make sure all supplies (fittings, capillaries, inline filters, columns, etc.) are
bio-inert / bio-compatible.
Be aware that some columns for bio-related applications have a stainless
steel case and can introduce iron and other metal ions in the flow path.
This may lead to adsorption of susceptible samples like phosphorylated
nucleotides. In this case, use PEEK-lined columns.
After using the system with high salt concentrations, flush it extensively with
water to prevent blockages caused by salt crystals.
Reliable operation of 1290 pumps during analysis cannot be guaranteed if
pressure falls below 20 bar. For optimal results, pressure should be at least
50 bar continuously. Therefore, when using columns that create low
backpressure (<50 bar, such as SEC columns with 1290 LC systems), install a
restriction capillary between the pump and the sampler, to achieve at least
50 bar.
Perform daily flush of the Multisampler with water if the Multiwash option is
installed (see Best Practices for Using an Agilent LC System Technical Note)
CAUTION
Agilent Bio-inert and Bio LC systems should not be subject to passivation or
similar procedures
This can cause irreversible damage to the system's internal surfaces
Do not perform passivation or similar procedures on bio-inert and
bio-compatible systems.

Table of Contents

Other manuals for Agilent Technologies 1290 Infinity II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 Infinity II and is the answer not in the manual?

Agilent Technologies 1290 Infinity II Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 Infinity II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals