EasyManua.ls Logo

Agilent Technologies 1290 Infinity II - Replace the Heat Exchanger (G7120 A)

Agilent Technologies 1290 Infinity II
291 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1290 Infinity II High-Speed Pumps User Manual 175
8Maintenance
Pump Head Maintenance (Tool Free)
Replace the Heat Exchanger (G7120A)
Tools required p/n Description
Wrench, 19 mm
5023-2501 Screwdriver Torx-T10
5067-5688 Torque wrench 1 25 Nm with 14 mm wrench
G4220-20013 4 mm hex bit
G4220-20015 Adapter ¼ in square to hex
G4220-20041 Bit Torx 10x25 mm
Parts required # p/n Description
1 G4220-81013 Heat Exchanger Channel A
(secondary pump head only)
OR 1 G4220-81012 Heat Exchanger Channel B
(secondary pump head only)
Preparations
Remove the pump head assembly from the pump
Remove the secondary pump head from the link plate
CAUTION
Loss of small spacer fitting
Inside the secondary pump head is a small spacer fitting, which can be
dropped easily when removing the heat exchanger.
The heat exchanger does not need to be removed for pump head
maintenance.
1 Remove the 19 mm screw at the front of the secondary
pump head.
2 Remove the front plate.