8Maintenance
Pump Head Maintenance (Tool Free)
3 Remove the screw at the bottom of the pump head. Do
not drop the golden spacer fitting.
4 Lift out the heat exchanger.
5 If removed, first insert the spacer fitting. Then insert the
new heat exchanger to the opening in the pump head
and lift it over the pins.
6 Use the 19 mm screw for fixing the front plate.
7 Insert and fix the screw.
Screw
Spacer fitting
Pins
Front plate
Screw
Screw
3 Nm
approx.