1290 Infinity II High-Speed Pumps User Manual 38
3Using the Pump
Best Practices
Best Practices
Daily / Weekly tasks
Daily Tasks
Replace solvents and solvent bottles for mobile phases based on
water/buffer.
Replace solvents and solvent bottles for organic mobile phase latest every
second day.
Check presence of seal wash solvent.
Purge each channel with fresh solvent at 2.5–3mL/min for 5min before
operation.
Equilibrate the system with composition and flow rate of subsequent method.
Weekly Tasks
Change seal wash solvent (10 % isopropanol in water) and bottle.
If applications with salts were used, flush all channels with water and remove
possible salt deposits manually.
Inspect solvent filters for dirt or blockages. Exchange if no flow is coming out
of the solvent line when removed from the degasser inlet.