1290 Infinity II High-Speed Pumps User Manual 44
3Using the Pump
Best Practices
Seal Wash
Seal Wash guarantees a maximum seal life time. Use Seal Wash:
When using buffers with elevated salt concentrations
When using volatile solvents with non-volatile additives
How to deal with solvents
Use clean bottles only.
Exchange water-based solvents daily.
Select solvent volume to be used up within 1 2 days.
Use only HPLC-grade solvents and water filtered through 0.2 µm filters.
Label bottles correctly with bottle content, and filling date / expiry date.
Use solvent inlet filters.
Reduce risk of algae growth: use brown bottles for aqueous solvents, avoid
direct sunlight.
CAUTION
Contaminated seal wash solvent
Do not recycle seal wash solvent to avoid contamination.
Weekly exchange seal wash solvent.