1290 Infinity II High-Speed Pumps User Manual 27
2 Site Requirements and Specifications
Performance Specifications
Performance Specifications
Performance Specifications Agilent 1290 Infinity II
High-Speed Pump (G7120A)
Table 3 Performance Specifications G7120A
Type Specification Comments
Hydraulic system Two dual pistons in series pump
with proprietary servo-con-
trolled variable stroke design
and smooth motion control for
active damping.
Pump resolution step size 300 pL
Flow range Settable: 0.001 5 mL/min, in
0.001 mL/min increments (exe-
cuted in 300 pL/step incre-
ments)
Flow precision 0.07 % RSD or 0.01 min SD,
whichever is greater
Based on retention time at con-
stant temperature
Flow accuracy ±1% or ±10 µL/min, whichever
is greater
Pressure operating range Up to 130 MPa (1300 bar) at
0–2mL/min
ramping down to 80 MPa
(800 bar) at 5 mL/min
Pressure pulsation <1 % amplitude or <0.5 MPa
(5 bar), whichever is greater
Compressibility compensation Automatic
Recommended pH range 1.0 12.5 Solvents with pH <2.3 should
not contain acids which attack
stainless steel
Gradient formation High-pressure binary mixing
Delay volume As low as 45 µL (10 µL without
mixer)
Composition range 0 100 %

Table of Contents

Other manuals for Agilent Technologies 1290 Infinity II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 Infinity II and is the answer not in the manual?

Agilent Technologies 1290 Infinity II Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 Infinity II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals