1290 Infinity II High-Speed Pumps User Manual 208
8Maintenance
Prepare the Pump Module for Transport
Prepare the Pump Module for Transport
When
If the module shall be transported or shipped.
Parts required p/n Description
9301-0411 Syringe; Plastic
9301-1337 Syringe adapter
G7120-44000 Protective Foam
Preparations
Flush both solvent channels with isopropanol, see Best Practices for Using an Agilent LC System
Technical Note.
Heavy weight
The module is heavy.
Carry the module at least with 2 people.
Avoid back strain or injury by following all precautions for lifting heavy
objects.
Ensure that the load is as close to your body as possible.
Ensure that you can cope with the weight of your load.
Mechanical damage
For shipping the module, insert the Protective Foam to protect the module
from mechanical damage.
Be careful not to damage tubing or capillary connections while inserting the
module in the Protective Foam.