1290 Infinity II High-Speed Pumps User Manual 208
8Maintenance
Prepare the Pump Module for Transport
Prepare the Pump Module for Transport
When
If the module shall be transported or shipped.
Parts required p/n Description
9301-0411 Syringe; Plastic
9301-1337 Syringe adapter
G7120-44000 Protective Foam
Preparations
Flush both solvent channels with isopropanol, see Best Practices for Using an Agilent LC System
Technical Note.
Heavy weight
The module is heavy.
 Carry the module at least with 2 people.
 Avoid back strain or injury by following all precautions for lifting heavy
objects.
 Ensure that the load is as close to your body as possible.
 Ensure that you can cope with the weight of your load.
Mechanical damage
 For shipping the module, insert the Protective Foam to protect the module
from mechanical damage.
 Be careful not to damage tubing or capillary connections while inserting the
module in the Protective Foam.