EasyManua.ls Logo

Agilent Technologies 1290 Infinity II - Writing the Purge Valve Tag Failed; Current of Primary Pump Drive too High

Agilent Technologies 1290 Infinity II
291 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1290 Infinity II High-Speed Pumps User Manual 128
7 Error Information
Pump Error Messages
Writing the purge valve tag failed
Error ID: 22431
Writing the purge valve tag failed.
Parameter: None
Current of primary pump drive too high
Error ID: 22433
The current of the primary pump drive is too high.
Parameter: 1 or 4 referring to pump drive.
Probable cause Suggested actions
1 Purge valve head tag defect.
Replace purge valve head.
2 Purge valve tag reader is defect.
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1 Blockage of flow path between primary
pump head and pressure sensor, e.g. of the
heat exchanger.
Check for blockages in flow path.
Please contact your Agilent service represen-
tative.
2 Primary pump drive is defect.
Please contact your Agilent service representa-
tive.

Table of Contents

Other manuals for Agilent Technologies 1290 Infinity II

Related product manuals