1290 Infinity II High-Speed Pumps User Manual 142
8Maintenance
Install Fittings and Capillaries
Install Fittings and Capillaries
1 Install fittings and capillaries.
2 Tighten fittings and capillaries.
WARNING
Solvent can spray under high pressure.
Observe appropriate safety procedures (for example, goggles, safety
gloves and protective clothing), when opening flow path.
CAUTION
Deformation of fittings and seals
Liquid drops under high pressure act like solid parts. Tightening connections
under high pressure can deform or destroy fittings and seals.
Never tighten flow connections under pressure.
NOTE
The lifetime of a fitting depends on how firmly it has been tightened; firm
tightening reduces the lifetime.
If fitting has been overtightened, replace it.