1290 Infinity II High-Speed Pumps User Manual 45
3Using the Pump
Purging the Pump
Purging the Pump
When the solvents have been exchanged or the pumping system has been turned
off for a certain time (for example, overnight) oxygen will re-diffuse into the
solvent channel between the solvent reservoir, vacuum degassing unit (when
available in the system) and the pump. Solvents containing volatile ingredients
will slightly lose these. Therefore purging of the pumping system is required
before starting an application.
1 Initiate a purge in the controlling software with a Purge flow set to 3 mL/min
per channel.
2 Flush all tubes with at least 15 mL of solvent.
Table 5 Choice of Priming Solvents for Different Purposes
Activity Solvent Comments
After an installation
When switching between reverse phase and
normal phase (both times)
Isopropanol
Isopropanol
Best solvent to flush air out of the system
Isopropanol is miscible with both normal phase
and reverse phase solvents.
After an installation Ethanol or Methanol Alternative to Isopropanol (second choice) if no Iso-
propanol is available
To clean the system when using buffers
After a solvent change
Bidistilled water
Bidistilled water
Best solvent to re-dissolve buffer crystals
Best solvent to re-dissolve buffer crystals
Before turning off system for an extended
period of time
50 % methanol or 50 %
isopropanol in water,
without additives
NOTE
The pump should never be used for priming/purging empty tubings (never let the
pump run dry). Use a syringe to draw enough solvent for completely filling the
tubings to the pump inlet before continuing to purge with the pump.

Table of Contents

Other manuals for Agilent Technologies 1290 Infinity II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 Infinity II and is the answer not in the manual?

Agilent Technologies 1290 Infinity II Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 Infinity II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals