1290 Infinity II High-Speed Pumps User Manual 120
7 Error Information
Pump Error Messages
Waste counter limit exceeded
Error ID: 22056
The counter for the waste volume has exceeded the limit, which has been set in
the user interface.
• Parameter: None
Flow rate limit exceeded
Error ID: 22064
The flow rate of the binary pump has exceeded the limit, while the pump runs in
pressure controlled mode, e.g. during a pressure test.
• Parameter: None
Probable cause Suggested actions
1 The waste container is full.
Empty waste container.
2 Inappropriate setting for waste counter.
• Reset waste counter.
• Adjust waste counter limit.
Probable cause Suggested actions
1 Leak
Check for leaks in the pump and flow path.
2 Bottle empty.
Fill solvent bottle.
3 Shutoff valve closed.
Open shutoff valve.
4 Drift of pressure sensor (unlikely for short
tests taking only minutes).
Replace pressure sensor.

Table of Contents

Other manuals for Agilent Technologies 1290 Infinity II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 Infinity II and is the answer not in the manual?

Agilent Technologies 1290 Infinity II Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 Infinity II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals