1290 Infinity II High-Speed Pumps User Manual 118
7 Error Information
Pump Error Messages
Pressure below lower limit
Error ID: 22015
The pressure has dropped below the lower limit.
• Parameter: None
Target pressure not reached for binary pump
degasser
Error ID: 22031
The target pressure of the binary pump degasser has not been reached within
the expected time.
• Parameter: Pressure in mbar
Probable cause Suggested actions
1 Leak
Check for leaks.
2 Bottle empty
Check bottle filling.
3 Wrong solvent (viscosity)
Check solvent.
4 Inappropriate setting
Check flow rate and lower pressure limit.
5 Column degradation
Replace column.
Probable cause Suggested actions
1 Condensation in degasser chamber due to
temperature fluctuation.
Equilibrate and restart module. Use Evacuation
Mode if necessary.
2 Degasser is defect.
Please contact your Agilent service representa-
tive.

Table of Contents

Other manuals for Agilent Technologies 1290 Infinity II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 Infinity II and is the answer not in the manual?

Agilent Technologies 1290 Infinity II Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 Infinity II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals