Agilent Technologies 1290 Infinity II - Binary Pump Shutdown During Analysis; Reading the Pump Encoder Tag Failed

Agilent Technologies 1290 Infinity II
291 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1290 Infinity II High-Speed Pumps User Manual 121
7 Error Information
Pump Error Messages
Binary pump shutdown during analysis
Error ID: 22065
The binary pump has been shut down by the control software or control module
during an analysis.
Parameter: 0 for off, 1 for standby.
Reading the pump encoder tag failed
Error ID: 22402
Reading the pump encoder tag has failed.
Parameter: 1 4 referring to pump drive
Probable cause Suggested actions
1 Pump has been shut down.
Restart pump.
Probable cause Suggested actions
1 Defect connection between encoder and
main board.
Please contact your Agilent service representa-
tive.
2 Missing or defect tag Defect connection
between tag and encoder.
Please contact your Agilent service representa-
tive.