1290 Infinity II High-Speed Pumps User Manual 130
7 Error Information
Pump Error Messages
Pump drive error
Error ID: 22438, 22439
The pump drive failed during calibration.
• Parameter: 1 – 4 referring to pump drive
Pump drive stroke blocked
Error ID: 22441
During initialization the pump defines the operation position of the pump drives
and therefore the pistons. First the pump drive moves backwards to find a
mechanical stop within the ball screw. Afterwards, pistons move forwards for
finding the maximum available stroke volume. These values are expected within
a pre-defined range. "Maximum stroke too short" means that the outer drive
position is too close. This can be caused by a drive initialization without pump
head or if the pump head has not been installed properly (screws are loose).
• Parameter: 1 – 4 referring to pump drive
Probable cause Suggested actions
1 Pump drive motor defect or mechanically
blocked.
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1 Wiper shifted
Please contact your Agilent service representa-
tive.
2 Pump head blocks piston movement
Replace, clean or repair pump head.
3 Pump drive motor is mechanically blocked.
Please contact your Agilent service representa-
tive.

Table of Contents

Other manuals for Agilent Technologies 1290 Infinity II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 Infinity II and is the answer not in the manual?

Agilent Technologies 1290 Infinity II Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 Infinity II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals