Pump Error Messages
Writing the pump encoder tag failed
Error ID: 22405
Writing the pump encoder tag has failed.
Parameter: 1 4 referring to pump drive
Pump drive blocked or encoder failed
Error ID: 22406
Pump drive blocked or encoder failed.
Parameter: None
Probable cause Suggested actions
1 Defect connection between encoder and
main board.
Please contact your Agilent service representa-
tive.
2 Defect tag.
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1 Blockage of system before pressure sensor.
Check for blockage of e.g. outlet valve filter frit,
purge valve, heat exchanger.
2 Drive encoder failed.
Please contact your Agilent service representa-
tive.