1290 Infinity II High-Speed Pumps User Manual 119
7 Error Information
Pump Error Messages
Degasser's pressure limit violation
Error ID: 22032
Pressure too far above the limit.
Solvent counter exceeded limit
Error ID: 22055
The counter for the solvent volume has exceeded the limit, which has been set in
the user interface.
Parameter:
• Without Solvent Selection Valve:
0 for channel A, 1 for channel B
• With Solvent Selection Valve:
2 for channel A1, 3 for channel B1, 4 for channel A2, 5 for channel B2
Probable cause Suggested actions
1 Leak in degasser chamber or degasser tub-
ing.
Please contact your Agilent service representa-
tive.
2 Defect vacuum pump.
Please contact your Agilent service representa-
tive.
3 Degasser chamber empty or connected to
air.
Block unused degasser channels.
Probable cause Suggested actions
1 No solvent present.
Refill solvent bottle.
2 Inappropriate setting.
Check solvent counter setting in user interface.