1290 Infinity II High-Speed Pumps User Manual 119
7 Error Information
Pump Error Messages
Degasser's pressure limit violation
Error ID: 22032
Pressure too far above the limit.
Solvent counter exceeded limit
Error ID: 22055
The counter for the solvent volume has exceeded the limit, which has been set in
the user interface.
Parameter:
• Without Solvent Selection Valve:
0 for channel A, 1 for channel B
• With Solvent Selection Valve:
2 for channel A1, 3 for channel B1, 4 for channel A2, 5 for channel B2
Probable cause Suggested actions
1 Leak in degasser chamber or degasser tub-
ing.
Please contact your Agilent service representa-
tive.
2 Defect vacuum pump.
Please contact your Agilent service representa-
tive.
3 Degasser chamber empty or connected to
air.
Block unused degasser channels.
Probable cause Suggested actions
1 No solvent present.
Refill solvent bottle.
2 Inappropriate setting.
Check solvent counter setting in user interface.

Table of Contents

Other manuals for Agilent Technologies 1290 Infinity II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 Infinity II and is the answer not in the manual?

Agilent Technologies 1290 Infinity II Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 Infinity II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals