1290 Infinity II High-Speed Pumps User Manual 137
8Maintenance
Introduction to Maintenance
Figure 14 The hydraulic path
Recommended Interval for Preventive Maintenance
The recommended interval for preventive maintenance is:
100 L (150 L for Long Life Technology) or 1 year (whichever comes first).
This recommendation is valid for LC instruments on which “typical” applications
are running.
A “typical” application can be characterized as follows:
pressure range 100 800 bar,
flow rates 0.5 3.5 mL/min,
typical solvents used in reversed phase LC.
Solvent
Jet Weaver
Valve
Degassing
unit
Pump head A
Pump head B
SSV
High-Speed Pump
Seal wash pump
Seal wash sensor
Pressure sensor
to Multisampler

Table of Contents