1290 Infinity II High-Speed Pumps User Manual 196
8Maintenance
Replace the Outlet Valve (G7120A)
3 Remove the heat exchanger capillary by pushing the
connector up and pulling it out of the valve.
NOTE
A gold seal between outlet valve and heat exchanger capil-
lary is used for a tight connection.
4 Unscrew the outlet valve with a 14 mm wrench (1) and
remove it (2).
5 Mount the pump head to the module. Do not fix the
screws at this stage!
6 Insert the outlet valve into the pump head (1). Using a
torque wrench, set approx. 10 Nm and close the outlet
valve (2).
7 Remove the pump head from the module again. 8 Position the entrance slit for the heat exchanger capil-
lary to face exactly to it, and then seat the heat
exchanger capillary back into the outlet valve by moving
it into the valve and pressing it down.
(1)
(2)
(1)
(2)
(1)
(2)
approx.
10 Nm
1.
2.